-25 %
Blepharoplasty Instruments Tray
$299.00
$399.00
- Stock: In Stock
- Article #: Blephset-002
Blepharoplasty Instruments Owing to the shortage of a typical nomenclature across various manufacturers, the outline of blepharoplasty surgical instrumentation remains nebulous among surgeons, surgical staff, and trainees of all levels. The instruments discussed below are found altogether surgical trays used for oculofacial cosmetic surgery or eye plastic surgery at the University of Iowa.
- Blepharoplasty instruments Tray Set consisting of the following tools.
Stevens Tenotomy Scissors, Blunt, Curved,11 cm, 4-1/2" 1
Iris Scissors, Flat Shanks, 10 cm, 4", Curved 1
Blepharoplasty Scissors, Sharp, Bevel Blades, Cvd, 4" 1
Kaye Blepharoplasty Rake, 5 Prong, Large, 13 cm, 5-1/4" 2
Castroviejo Suture Forceps, 10.5 cm, 4-1/4", 5 mm Tip 1
Castroviejo Suture Forceps, 10.5 cm, 4-1/4", 9 mm Tip 1
Total………………….23
- Sterilization: Surgical instruments may be steam autoclaved at a temperature of 121°C (250°F) for 30 minutes or flash autoclaved at a temperature of 127° to 132°C (260° to 270°F) for 10 minutes.
- Recommended Sterilization Parameters: PRE-VAC Unwrapped Goods Cycle 5 minutes at 132° C/270° F. Dry Time: none recommended.
Wrapped Goods Cycle 5 minutes @ 132° C/270° F Dry Time: 5‐20 minutes*
- Recommended Gravity. Unwrapped Goods Cycle 10 minutes @ 132° C/270° F Dry time: none recommended. Wrapped Goods Cycle 30 Minutes @ 132° C/270° F Dry Time: per facility recommendation. Wrapped Goods Cycle 60 Minutes @ 132° C/270° F Dry Time: per facility recommendation. *Depending on Hospital Protocol* Warning: Flash sterilization is not the recommended sterilization method. All surgical instruments and parts should be steam autoclaved when time permits. However, flash sterilization may be used when there is an immediate need for an instrument, injector, or accessory and there is no other alternative.
- Aeration: As required to return to room temperature or to start the next cycle.
- Contraindication: Hospital policy and procedure(s) take precedence over this protocol. Failure to use approved cleaning and sterilization procedures may void the warranty.
- Cleaning Instructions: Cleaning: All instruments must be cleaned to remove contaminated substances before sterilization. After each use, perform an initial cleaning within the sterile field by wiping the tool or instrument free of blood and body fluids. Then clean the instrument thoroughly with a mild soapy solution. Do not use alkaline-based cleaners. Complete preparation for sterilization by hospital procedure using an approved cleaning agent. Do not run any type of abrasive materials or steel brushes through the lumen of the tool. The abrasive material can score the inside of the tool, rendering them inefficient. After flushing, the instrument needs to be well-rinsed with warm to hot water. Visually inspect each instrument to ensure all fatty deposits are removed. All instruments should be cleaned and rinsed with mild, non-abrasive, grease-removing cleaners before they are sterilized. Gulmaher Instruments recommends that during the cleaning and rinsing procedures, all instruments should be inspected. Ultrasonic cleaning after manual cleaning is recommended.