Care & Sterilization of Surgical Instruments

These guidelines apply to reusable stainless steel instruments supplied by Gulmaher Surgico. Always follow your facility's validated reprocessing procedures.

Before first use

Instruments are supplied non-sterile. Remove all packaging, then clean and sterilize every instrument before its first use.

Cleaning after use

  • Clean instruments as soon as possible after use. Do not let blood or tissue dry on the surface.
  • Open hinged instruments and disassemble multi-part instruments before cleaning.
  • Use a neutral or mild alkaline, enzymatic detergent. Clean manually with a soft brush or in an ultrasonic cleaner.
  • Rinse thoroughly, preferably with distilled or deionized water, and dry completely.

Inspection and lubrication

  • Check each instrument for damage, corrosion, loose parts and correct function before sterilizing.
  • Lubricate hinges, box locks and moving parts with a water-soluble instrument lubricant.

Sterilization

Stainless steel instruments are suitable for steam sterilization (autoclave). Use a validated cycle in line with your facility's procedures and recognized standards such as ISO 17664 and AAMI ST79.

What to avoid

  • Saline, chlorine or bleach solutions, which cause pitting and corrosion
  • Steel wool, wire brushes and abrasive cleaners
  • Mixing instruments of different metals in the same cleaning cycle

Storage

Store clean, dry instruments in a clean, dry area, protected from impact.